Further analysers available
|
 |
|
- Ammonia (NH3)
|
|
- Chloride (Cl2)
|
|
- Fluoride (Fl)
|
|
- Iodide (I)
|
|
- Sulfide (S-)
|
|
- General technical information
|
 |
 |
|
Ammonia (NH3)
|
 |
|
Measuring range
| |
0.005 ... 2.0 mg/l NH3 (low range)
|
|
| |
0.5 ... 50.0 mg/l NH3 (high range)
|
|
Resolution
| |
0.01
|
|
Accuracy
| |
+/- 2% (based on full scale)
|
|
Measuring interval
| |
Continuous
|
|
| |
Discontinuous/programmable
|
|
| |
External start
|
|
Measuring method
| |
Ion selective electrode
|
|
Reagents
| |
2
|
|
Calibration
| |
Automated
|
|
Cleaning
| |
Automated
|
|
Interferences
| |
Hg, Cu, Zn, Cd, Co, Ni, Au, Ag
|
|
|
|
Interfering ions can be masked in local applications.
|
|
Top of page
|
 |
|
Chloride (Cl2)
|
 |
|
Measuring range
| |
0.1 ... 10.0 g/l Cl
|
|
Resolution
| |
0.01
|
|
Accuracy
| |
+/- 3% (based on full scale)
|
|
Measuring interval
| |
Continuous
|
|
| |
Discontinuous/programmable
|
|
| |
External start
|
|
Measuring method
| |
Ion selective electrode
|
|
Reagents
| |
3
|
|
Calibration
| |
Automated
|
|
Cleaning
| |
Automated
|
|
Interferences
| |
Bi, Cd, Mn, Pb, Sn, Tl
|
|
|
|
Interfering ions can be masked in local applications.
|
|
Top of page
|
 |
|
Fluoride (Fl)
|
 |
|
Measuring range
| |
0.2 ... 2000 mg/l F
|
|
Resolution
| |
0.1
|
|
Accuracy
| |
+/- 2% (based on full scale)
|
|
Measuring interval
| |
Continuous
|
|
| |
Discontinuous/programmable
|
|
| |
External start
|
|
Measuring method
| |
Ion selective electrode
|
|
Reagents
| |
1
|
|
Calibration
| |
Automated
|
|
Cleaning
| |
Automated
|
|
Interferences
| |
OH=0.1
|
|
|
|
Interfering ions can be masked in local applications.
|
|
Top of page
|
 |
|
Iodide (I)
|
 |
|
Measuring range
| |
0.005 .. 1.0 mg/l I-
|
|
Resolution
| |
0.01
|
|
Accuracy
| |
+/- 3% (based on full scale)
|
|
Measuring interval
| |
Continuous
|
|
| |
Discontinuous/programmable
|
|
| |
External start
|
|
Measuring method
| |
Ion selective electrode
|
|
Reagents
| |
2
|
|
Calibration
| |
Automated
|
|
Cleaning
| |
Automated
|
|
Interferences
| |
Cl, Br, S-, CN, S2O3
|
|
|
|
Interfering ions can be masked in local applications.
|
|
Top of page
|
 |
|
Sulfide (S-)
|
 |
|
Measuring range
| |
0.02 ... 50.0 mg/l S-
|
|
Resolution
| |
0.1
|
|
Accuracy
| |
+/- 3% (based on full scale)
|
|
Measuring interval
| |
Continuous
|
|
| |
Discontinuous/programmable
|
|
| |
External start
|
|
Measuring method
| |
Colourimetric
|
|
Reagents
| |
2
|
|
Calibration
| |
Automated
|
|
Cleaning
| |
Automated
|
|
Interferences
| |
Hg, Cu, Zn, Cd, Co, Ni, Au, Ag
|
|
|
|
Interfering ions can be masked in local applications.
|
|
Top of page
|
 |
|
General technical information
|
|
|
Power supply
|
|
230 VAC, 50 Hz (110 VAC optional)
|
|
Current consumption
|
|
Max. 50 VA
|
|
Switch contacts
|
|
2 x P-controlled (min, max)
|
|
|
|
1 x alarm contact
|
|
Analogue output
|
|
4..20 mA
|
|
Digital output
|
|
RS232 C
|
|
|
|
Options: Interbus-S, Profibus, CAN
|
|
Cabinet size
|
|
746 x 600 x 373 mm (HxBxD)
|
|
Weight
|
|
App. 80 kg
|
|
Environmental temperature
|
|
+15°C .. +35°C
|
|
Humidity
|
|
Max. 70% RH, no condensation
|
|
Sample pressure
|
|
0 .. 0.5 bar
|
|
Sample inlet
|
|
6/4 mm hose
|
|
Sample drain
|
|
Unpressurised, free falling, 6/4 mm hose
|
|
 |
Top of page
|
SEIBOLD-Messtechnik und Projektentwicklug GmbH, A-1110 Wien, Grillgasse 51
Tel +43 1 332 00 60, Fax +43 1 332 00 60 44,
office@seiboldme.com
|
|
|